| STEP | ACTION |
|---|---|
| 1 | Use Device Explorer to turn the BCPM manager off then back on. |
| 2 | Run a short monitor scan to confirm the BCPM is operational. If not, go to step 3. |
| 3 | Quit and relaunch the BCPM screens. After they've launched, turn the BCPM manager off then on again. |
| 4 | Run a short monitor scan to confirm the BCPM is operational. If not, the BCPM control computer or rack may have a problem. Call for assistance. |
| STEP | ACTION |
|---|---|
| 1 | Confirm the BCPM hardware is powered ON. If off, use the other recovery procedure. |
| 2 | Turn the BCPM control computer ON using the power button on the back of the computer case. |
| 3 | Check the Cleo message window for problems. |
| 4 | Confirm BCPM setups prior to resuming observing. |
| STEP | ACTION |
|---|---|
| 1 | Power up BCPM hardware as follows: Turn ON the top power switch (back of BCPM, top rack). Next, turn ON the bottom power switch (also in back of BCPM, bottom most rack). |
| 2 | Turn the BCPM control computer OFF (if not already) by logging into the BCPM2 terminal as root then using the following command: sync; ^ init ^ 5 (Wait for the computer to power down) |
| 3 | Turn the BCPM control computer ON using the power button on the back of the computer case. |
| 4 | Check the Cleo message window for problems. |
| 5 | Confirm BCPM setups prior to resuming observing. |